GLUCOSE BIOSENSOR -
METHOD OF OPERATION
The screen display provides a menu to follow for
each operation. The keys that are available for operation are
shown on the right hand side of the screen display during each
step.
Important keys are:
| Dose |
Prompts the probe to exude paste from the cartridge. |
| Calibrate |
Enters the calibration menu. |
| Parameter |
Allows the measurement conditions and display options
to be altered. This also allows for changing and
numbering of cartridges. |
| Measurement |
This initialises the 20 second measurement period. |
| Stop |
Use as an escape key.
|
To insert a new cartridge
- Choose the Parameter option
- Q-Change cartridge? A-Y
- Enter the cartridge number.
- Enter the cartridge type G2
Follow the cleaning procedure outlined on the screen to
prepare the electrode surface.
To Calibrate
- Choose the Calibrate option.
- Choose 'Calibrate with multiple standard solutions'.
- Enter concentrations and follow the cleaning and
measurement procedure outlined, using first the 0.5M then
the 1.0M and finally the 2.0M glucose solutions. Pipette
1ml of each solution into a 2ml vial, place into the
centre hole on the stand and immerse the probe tip fully.
The probe tip should be wiped clean with a tissue between
measurements to prevent contamination of subsequent
solutions.
The calibration procedure leads directly into the measurement
procedure, allowing for the analysis of unknown solutions.
Calibration and measurement must be performed at the same
temperature as we are using an enzyme reaction.
Calibration should be performed on each new cartridge and
before each series of analysis.
Analysis of Potatoes
- Peel and juice potatoes.
- Allow Potato juice to settle.
- Insert sensor cartridge (See procedure above).
- Calibrate the cartridge (See procedure above).
- Check that a correlation coefficient of at least 0.990 is
achieved. If this is not achieved the calibration
procedure should be repeated.
- To ensure that standard conditions of pH and ionic
strength are employed during both measurement and
calibration, the potato juice must be added to a buffer
solution. This can be completed by
1) Accurately weighing equal masses of potato juice and
0.2M buffer solution into a vessel and taking 1ml
aliquots for measurement, or
2) Pippeting 0.5ml of potato juice and 0.5ml of 0.2M
buffer into a 2ml vial.
- Prepare solutions for testing.
- Run as many analysis as are required, following the
screen prompts to Dose and clean
the probe tip.
- Do not leave the probe tip immersed in the test solution
for longer than necessary.
- Wipe clean the probe tip with tissue between each
measurement to prevent contamination of subsequent test
solutions.
- When prompted for 'Dilution Factor' enter a factor of 2.
This allows for the addition of buffer solution outlined
above.
- The screen will display the glucose concentration in the
chosen units.
To achieve the best results each test should be performed in
the same routine, using a similar pressure when cleaning the
probe tip.
Back to Information
© Gwent Sensors Ltd 2008
These pages were last updated on 3rd April 2008